2.4 Setting the RS-232 Communication Parameters

2.4.1

Default RS-232 Communication Parameters.

The Default RS-232 Communication parameters are shown in the

table below.

Baud rate

9600

Stop bits

1

Data bits

7

Parity

Even

Hardwire mode

Leased line

Handshake type

Hardwire

Text Line

Terminator

CR-LF

Performing a full reset will set all parameters to the default settings.

2.4.2

Checking/Changing the RS-232 Communication

Parameters

The communication parameters for the serial interface must be set

using the push-buttons on the front of the monitor.

1. Press SET-UP S3 S1 S3 S1 .The screen display now shows the

following text.

If the baud rate displayed is incorrect press

¿

and use ▲ and ▼ to

display the correct value. Press

¿

again to store the selection.

If the baud rate displayed is correct, then press ▼ to continue to the

next parameter.

Press S1 to select 1 STOP BIT.

Press S1 to select 7 DATA BITS

Press S2 to select EVEN PARITY

Press S3 to select LEASED-LINE

Press S3 to select HARD-WIRED HANDSHAKE.

Press SET-UP to exit the set-up mode

Press RESET and S1 in order that the new settings are enabled.