When the installed optical filter has been zero-point, humidity-

interference and span calibrated then the Monitor is able to perform

accurate measurements.

Fig.12.3 Part of a calibration data print-out

12.3.3 Checking and Changing Optical Filter Parameters

Before starting calibration it is always wise to check that information

about the installed optical filter has been “entered” correctly in the

active set-up of the Monitor. If the incorrect UA number is

“entered” for the optical filter, the incorrect optical filter factors will

be used in the calculation of the calibration factors, and calibration

will therefore be useless.

The UA number can only be changed using the push-buttons on the

front panel of the monitor.