Gas concentrations – two-point calibration, Simple test to determine the linearity – LumaSense Technologies 3434i SF6 Leak Detector User Manual

Page 163

Advertising
background image

Chapter 12

______________________________________________________________________

_____________________________________________________________________________

BE6028-14

3434i SF6 Leak Detector

LumaSense Technologies A/S

Page 163 of 191

Gas Concentrations – Two-point Calibration

During a two-point span calibration, the total signal in the cell (V

low

)

is measured first with a low concentration (C

low

) of span gas, and

then the total signal in the cell (V

high

) is measured with a high

concentration (C

high

) of span gas. This enables the Monitor to

compensate for any un-linearity in the relationship between the

signal measured in the cell and the concentration of the gas in the

cell (this is illustrated, for a gas A1, in

Fig.12.14a

).

Here are some guidelines about the concentration of gas you should

use during a two-point span calibration:

·

The Low Concentration should be 10 to 30% of the highest

concentration you expect to measure.

If the desired measurement range is mainly within the linear

range then the low concentration should be between 1000 and 10

000 times the detection limit.

If the desired measurement range is mainly above the linear

range then the low concentration should be more than 10 000

times the detection limit.

·

The High Concentration should be 75% to 100% of the highest

concentration you expect to measure, but no higher than this.

Simple Test to Determine the Linearity

Let us assume that you wish to measure a gas over a dynamic range

from 100 ppm and 3000 ppm of a gas that has a detection limit of

0.3 ppm.

In this case, the range in which you want to measure is from 333

times to 10 000 times the detection limit. With our general

guidelines, you’re not able to decide whether you have to perform a

single- or two-point calibration. Therefore, we advise you to perform

the following test:

1. Perform a single-point calibration with a low concentration of the

gas. Use 10% of the highest expected concentration level. In this

case, use 300 ppm (1000 times the detection limit).

2. Perform a measurement with a high concentration of the gas.

Use, for example, 90% of the highest expected concentration

level. In this case, use 2700 ppm.

If the measured gas concentration differs by more than 5% from the

real concentration, then a two-point calibration with 2 gas

concentrations is recommended.

Advertising