7 restoring of calibration data – LumaSense Technologies 3434i SF6 Leak Detector User Manual

Page 25

Advertising
background image

Chapter 2

______________________________________________________________________

_____________________________________________________________________________

BE6028-14

3434i SF6 Leak Detector

LumaSense Technologies A/S

Page 25 of 191

The following text is displayed.

1. Press S1 “CHANGE USER LEVEL”

The following text appears.


2. Press

¿

.

3. Enter the current Password using the andkeys.

4. Press

¿

.

The following text appears.


5. Press

¿

.

6. Use the ▲ and ▼ keys to change the User level.

7. Press

¿

.

8. Press the Lock function push button to exit.

2.7 Restoring of calibration data

If you have ordered a calibration from our calibration laboratory, you

will receive a CD with a backup of the calibration data. It is very

important that you store these data in your Calibration Software

BZ7002. Please store the data before using the Gas Monitor. Please

refer to the User Manual BE6034 regarding this procedure.

ENTER PASSWORD "*****"

PRESS ENTER TO CHANGE VALUE

CURRENT USER LEVEL IS X

CHANGE USER LEVEL

CHANGE PASSWORD

CURRENT USER LEVEL IS X

PRESS ENTER TO CHANGE VALUE