The maximum time you can set a pre-set monitoring period to be is

7 days. If you try to enter a greater number of days, the Monitor will

automatically change the period to 7 days. However, if you do not

set a pre-defined monitoring period, the Monitor will just store as

much data as possible before starting to overwrite the “oldest” stored

measurement results.

4.2 Attaching the Sampling Tube and External Filter

1 Cut a short section of sampling tube of Poly-Tetra-Fluoro-

Ethylene (Teflon) tubing, (optional accessory).

2 Using Fig.4.3 and the instructions below, attach one end of this

tube to the air-inlet stub on the back-panel of the Monitor.

Push one end of the Teflon tubing through the non-threaded end

of the nut.

Hold the end of the tubing between the fingers and gently push

the tubing over the end of the Monitor’s air-inlet stub as far as it

will go.

Note: if the tubing is bent/broken during this procedure, remove

the tubing from the stub and repeat this step using an

undamaged length of tubing.

Screw the threaded-nut firmly onto the end of the air-inlet stub.

Fig.4.3 Attaching tubing to the air-inlet stub and the external filter