Alarm set-up – LumaSense Technologies 3434i SF6 Leak Detector User Manual

Page 80

Advertising
background image

Chapter 7

______________________________________________________________________

_____________________________________________________________________________

BE6028-14

3434i SF6 Leak Detector

LumaSense Technologies A/S

Page 80 of 191

Table 7.6 The range of acceptable values and the default value of the

averaging period

Display Text

Acceptable

Values

Default Value

ENTER AVERAGING PERIOD

00:10

hours:minutes

00:10 to 12:00

hours:minutes

00:10

Stand-alone Use

When the averaging function is active it will not only affect the

results displayed on the screen but also the measurement results

which are being printed out from the Display Memory (see

Chapter

10

). However, when the averaging function is deactivated – by

pressing AVERAGE again (the LED will switch off) – the original

results (un-shaded) will replace the averaged results on the screen.

Gas :

Gas Setup: there are up to six selections here: A,B, C, D, E and W.

These represent the filter positions in the monitor’s filter carousel.

When setting up a new monitoring task, you can choose freely

between the gases available for the filter (names entered when

calibrating the filter, see sections

12.3

and

12.5

). Once a gas has

been selected for the filter and the measurement task started, you

can no longer change the gas for this filter in this measurement task.

However, you can still select gases for filters, which have not

previously been used in the measurement task.

Note: only one gas per filter can be selected at a time for

measuring. The W position always represents the water filter

installed in the monitor.

Sample Integration Time: this enables you to make a choice

between speed and accuracy. Generally, the longer the integration

time the more accurate the results. There are several selection

possibilities here:

Integration Type

Integration Time

Fast

1 s

Normal

5 s

Low Noise

20 s

Advanced

0.5s, 1s, 2s, 5s, 10s, 20s, 50s

Alarm set-up:

Select Gas: this option enables you to select any of the gases (one

per filter position) already defined in the set-up.