12.9.2 Performing a Zero-point Calibration of the Water-

vapour Filter

As explained in the previous section, a zero-point calibration of the

water-vapour filter has to be done together with the zero-point

calibration of the other installed optical filters. The latter task has

been fully described in

section 12.6

. It is worth-while to emphasise

again that it is vitally important to use perfectly dry zero-gas (that

is, zero-gas which contains 0% water-vapour) during this calibration.

Note that during the zero-point calibration of filter “A” (task 1,

shown below), and during the combined zero-point calibration of

filter “A” and the water-vapour filter (task 2, shown below), a zero-

gas supply is attached to the air-inlet of the Monitor. The Monitor

measures the signal in the cell using the water-vapour filter and the

other installed optical filter. Signals measured with filter “A” enable

the concentration offset factors for filter “A” to be stored in the

filter bank which is active during calibration. However, the signal

measured with the water-vapour filter is used differently:

In Task 1:

The signal measured with the water-vapour filter is used to

compensate all measured signals for any interference caused by the

presence of any residual water-vapour in the zero-gas supply.

In Task 2:

The signal measured with the water-vapour filter is used to update

the concentration offset factor for the water-vapour filter. This

means that the signals measured with filter “A” be compensated for

the presence of any water vapour present. This is why it is vitally

important to use perfectly dry zero-gas during this task.

Please refer to

section 12.6.2

for details about how to perform a

zero-point calibration task.

12.9.3 Setting-up a Span Calibration of the Water-vapour

Filter

Remember that span calibration of the water-vapour filter is only

necessary if you wish to measure the absolute concentration of water

vapour in air samples. If the water-vapour filter is not span

calibrated you will not be able to measure the concentration of water

vapour in gas samples during a monitoring task.

Before the water-vapour filter can be span calibrated, it has to be

zero-point calibrated, that is, a concentration offset factor for the