LumaSense Technologies 3434i SF6 Leak Detector User Manual

Page 184

Advertising
background image

Chapter 13

______________________________________________________________________

_____________________________________________________________________________

BE6028-14

3434i SF6 Leak Detector

LumaSense Technologies A/S

Page 184 of 191

SIGNIFICANT CHANGE IN CALIB. FACTORS

This message is associated with the span calibration of one (or more)

of the filters.

The above message indicates that the Conversion factor calculated

during span calibration of a particular filter deviates more than 15%

from the Conversion factor calculated during the last span

calibration of this filter. If you are absolutely sure that the previous

span calibration of this filter was correctly performed then this

message could indicate:

1. That you have “entered” the wrong concentration for the span-gas

in the span calibration set-up.

2. That the concentration of your span gas is not correctly quoted on

its “Analysis Certificate”.

3. That there is a leak in the tubing/fittings in your calibration

equipment, which alters the concentration of the span gas

reaching the cell.

HIGH CONC. SPAN MEASUREMENT TOO LOW

This message is associated with a two-point span calibration task.

The above message indicates that the signal measured during span

calibration, when the high concentration of span gas is in the cell, is

less than 15µV. This means that the “high” concentration of span gas

is too low.

LOW CONC. SPAN MEASUREMENT TOO LOW

This message is associated with a two-point span calibration task.

The above message indicates that the signal measured during span

calibration, when the low concentration of span gas is in the cell, is

less than 3µV. This means that the “low” concentration of span gas is

too low.