Choosing a filter bank – LumaSense Technologies 3434i SF6 Leak Detector User Manual

Page 160

Advertising
background image

Chapter 12

______________________________________________________________________

_____________________________________________________________________________

BE6028-14

3434i SF6 Leak Detector

LumaSense Technologies A/S

Page 160 of 191

During span calibration a supply of a particular gas (e.g. gas A1) of

known concentration is attached to the air-inlet of the Monitor and

the total signal in the cell is measured using the water-vapour filter

and the filter which is being span calibrated (for example “A”). As

filter “A” has already been calibrated for humidity interference, the

total signal (V

total

) measured with filter “A” can be compensated for

any signal produced by water-vapour’s absorption of light from filter

“A” (V

H2O

) during the span calibration task. This means that your

span gas does not have to be perfectly dry. As the filter has already

been zero-point calibrated, the cell noise V

offset

(when filter “A” is

used) is known, and therefore the span calibration curve can be

drawn (see Fig.12.13).

Fig.12.13 Curve showing a linear span-calibration curve

Choosing a filter bank

If a filter (e.g. “A”) has already been calibrated to measure a gas

(e.g. A1) and you wish to calibrate with gas A1 again, then select the

filter bank containing the gas conversion factor for gas A1. Online

users can do this when downloading the calibration factors. Stand-

alone users should do this before setting-up and performing the span

calibration task. The gas conversion factor calculated during the

span calibration will then just overwrite the gas conversion factor

calculated during the previous span calibration of the filter with this

gas.

If you wish to span calibrate a filter to measure more than one gas,

then select a new filter bank each time you span calibrate with a

Advertising