Chapter 9

______________________________________________________________________

_____________________________________________________________________________

BE6028-14

3434i SF6 Leak Detector

LumaSense Technologies A/S

Page 108 of 191

Fig.9.3 Effect of changing a UA number in the “active” set-up

For example, if the UA number of the filter installed in position “A” of

the filter carousel is changed in the active set-up then calibration

factors are deleted from the active file “A” in the Working Memory

and from files A 1, A 2, A 3, A 4 and A 5 in the Source Memory of

the monitor (see Fig.9.3). The Monitor will not be able to measure

any gas with filter “A” until it has been fully calibrated, see

Chapter

12

for details.

9.1.6

Effect of a Full Reset on the Monitor’s Memory

A FULL RESET of the Monitor changes the data stored in its

Working Memory as follows:

1. All data stored in Display Memory and Background Memory is

deleted.

2. Values of the Optical Filter Set-up Parameters in Block 2 of The

Source Memory are copied into Block 2 of the Working Memory.

3. The active values of the other set-up parameters in Block 3 of the

Working Memory are replaced by the factory-set values defined

in Block 3 of the Source Memory.