3 tasks before starting any calibration – LumaSense Technologies 3434i SF6 Leak Detector User Manual

Page 138

Advertising
background image

Chapter 12

______________________________________________________________________

_____________________________________________________________________________

BE6028-14

3434i SF6 Leak Detector

LumaSense Technologies A/S

Page 138 of 191

Monitor. The signal in the cell is measured with the water vapour

filter and with the installed optical filters.

The water vapour in the cell absorbs light from the water-vapour

filter and from the installed filter, producing signals which are related

to water-vapour’s relative absorption of the light at the wavelengths

transmitted by the optical filter (see

Fig.12.2

). We know, however,

that the ratio of the signal measured with any one filter (e.g. “A”)

and the signal measured with the water-vapour filter is a constant.

This ratio is related to the Humidity Gain Factor calculated during

the humidity-interference calibration of filter “A”. For example, if the

signal produced by water vapour when using the water-vapour filter,

is 160µV and using Filter “A” is 16µV, then, if the signal measured

with the water-vapour filter is found to be 80µV, we know that this

concentration of water vapour will produce a signal of 8µV when

Filter “A” is used. Measurements made during humidity-interference

calibration of the installed filters therefore enable the Monitor to

compensate any measured signal for water-vapour’s interference.

12.3

Tasks Before Starting Any Calibration

Before starting any kind of calibration task there are several

operations, which need to be performed:

12.3.1 Obtaining a Print-out of Calibration Data in the

Monitor

Whether the Monitor is to be used online or as a stand-alone

instrument, the Calibration Software BZ7002 can be used to check

the Monitor’s existing calibration. Please refer to User manual

BE6034.

Stand-alone Use

With the monitor connected to a printer (details are given in

Chapter

10

):

Press PRINT and then S3.

To stop printing, press PRINT again.

12.3.2 Checking the Calibration of Each Installed Optical

Filter

A print-out of the calibration data in the Monitor’s memory is useful

to have for reference. It also enables you to find out how “old”

Advertising