VF Transmitter Menu Tree Software Release 04.02.XX

Level 1

Level 2

Level 3

Range

Description

Output Config

4-20mA
Ch1& Ch2

Output Sel

Flow Rate, Flow
Quality, Blank

Parameter to output

Power Sel

Internal, External

Power supply for 4-
20mA current

Low End

Based on ‘output sel’

Low end output (4mA)

High End

Based on ‘output sel’

High end output (20mA)

Out of Range

Hold, >20mA, <4mA,
4mA

Behavior when meter
reading is out of range
or meter has no reading

Overrange
Rail

Enable, Disable

‘Enable’ causes the
output to go to full
(20mA) or min (4mA)
when system over
ranges

4mA Trim

2 to 6

Adjust 4mA output

20mA Trim

18 to 22

Adjust 20mA output

Pulse

Multiplier

0 - 999999

Multiplier on output

Width (ms)

.5,1,20,33,50,100

Pulse width

Lowcut

0% to 100%

Low end cutoff

Output Sel

Flow Rate, Flow
Quality, Totalizer, Flow
Rate%

Parameter to output

Alarm
Control

Warning

Off/On/Equation Input

Activates alarm warning
function

Critical

Off/On/Equation Input

Activates alarm critical
function

Manual Clear Enable / Disable

Allows for manual or
automatic alarm
clearing

Alarm
Warning
Threshold

Flw Min

0-100% of range

Used to set the
minimum and maximum
values for a warning
alarm

Flw Max

Alarm Critical
Threshold

Flw Min

0-100% of range

Used to set the
minimum and maximum
values for a critical
alarm

Flw Max

Flow
Damping
Filter

State

Enable, Disable

Enable Damping

Time
Constant (s)

0-600 seconds; default
is 3

Smoothes output due to
rapid changes in flow

Flow Noise
Filter

State

Enable, Disable

Enable noise filtering

Magnitude

Low, High

Amount of damping