Gvf meter faults – CiDRA SONARtrac PW VF-100 User Manual

Page 272

Advertising
background image

20838-01 Rev 09

Page J-13

GVF Meter Faults:

Item
No

PROBLEM

POSSIBLE
CAUSE

DIAGNOSTIC
METHOD

DIAGNOSTIC RESULT

COMMENTS

19

The GVF
Meter
Displays
dashes “-----“

The acoustics at the
measurement location
are too low

Contact Technical
Support.

20

The GVF
Reading is
erratic

Signals may be
saturated

Select and run
‘AUTOSET GAIN’
found in GAIN
submenu of
DIAGNOSTICS
menu

Yes

– this corrected the problem

Acquire Snapshot, Upload
to secure Website, and
contact Customer Support.

No

– the problem persists

Acquire Snapshot, Upload
to secure Website, and
contact Customer Support.

21

The GVF
Meter always
reads 50%
GVF or some
other value
that is not
correct

Transmitter setup
parameters are
incorrect

Verify setup
parameters to ensure
they are correct

Yes

– the setup parameters are correct

The actual GVF of the fluid
is >50% or what is
indicated on the transmitter

Yes

– the setup parameters are correct

but the GVF is definitely incorrect

Acquire Snapshot, Upload
to secure Website, and
contact Customer Support.

No

– the setup parameters are not correct

Correct transmitter input
parameters in BASIC
CONFIG menu