4 alarm warn threshold and alarm crit threshold, Table 2, Alarm triggers -15 – CiDRA SONARtrac PW VF-100 User Manual

Page 159

Advertising
background image

20834-01 Rev 04

Page 8-15

8.3.4

Alarm Warn Threshold and Alarm Crit Threshold

These menu selections are used to set the flow rate and gas volume
fraction alarm set points. All other parameters must be set using the
Configure menu options in the passive sonar meter software utilities or
by a configuration file change using the USB Port.

The following table lists the alarm parameters with their default values.

Alarm Parameter

Condition

User

Entry

Warning

Default Value

Critical

Default Value

Band Temperature

TMP>

Y

> 80 °C

> 90 °C

TMP<

Y

< 0 °C

< 0 °C

RMS Sound Pressure
Level

SPL>

Y

> 200 dB

> 200 dB

SPL<

Y

< 50 dB

< 80 dB

Volumetric Flow Quality

VFQ<

Y

< 0.3

< 0.2

Speed Of Sound Quality

SSQ<

Y

< 0.03

< 0.01

Event log updated

LOG

N

Sensor overload

OVL

N

Volumetric Flow or True
Liquid Flow

FLW>

Y

> 30 ft/s

> 30 ft/s

FLW<

Y

< 3.2 ft/s

< 3 ft/s

Gas Volume Fraction

GVF>

Y

> 100 %

> 100%

GVF<

Y

< 0%

< 0%

Sensor Failure

FAIL

N

Table 2

Alarm Triggers

Note: The alarm trigger for Flow (VF or TLF) and Gas Volume Fraction
(GVF) can be set from the transmitter menu. All other values must be
set using passive sonar meter software utility or by configuration file
change through the USB Port.

Up to three alarm triggers can be selected. For example, an alarm
output may be programmed like this:

Warning = TMP I LOG I VF

(Warning = Band Temperature or Event Log Updated or VF)

Critical = FAIL I OVL & LOG

(Critical = Sensor Fail or Sensor Overload and Event Log Updated