System faults (page 3) – CiDRA SONARtrac PW VF-100 User Manual

Page 263

Advertising
background image

20838-01 Rev 09

Page J-4

System Faults (page 3):

Item
No

PROBLEM

POSSIBLE
CAUSE

DIAGNOSTIC
METHOD

DIAGNOSTIC
RESULT

COMMENTS

6

Screen is
light or dark
in color

Incorrect screen
contrast setting

Check screen setting in
‘DISPLAY /
CONTRAST in
CUSTOMIZE menu.
Normal setting is 170

No

– the contrast setting

is not 170

Adjust contrast setting up or down as
necessary. If problem persists contact
Customer Support.

Yes

– contrast is set to

170

Try adjusting contrast. If this does not work
contact Customer Support.

7

The 4

–20mA

input on plant
data system
does not
work

Incorrect wiring

Verify 4-20mA is wired
to proper connector
terminals in the
transmitter (internal or
external power)

No

– system is not

connected to
appropriate terminals

Reconnect wires; refer to section 8 of
Transmitter Installation Manual

yes

See below

Verify ‘POWER SEL’
on ‘4 -20mA CH1’ (or 2)
in OUTPUT CONFIG
menu (internal or
external) is set to
correspond with the
wiring option above

No

– incorrect ‘Power

Sel’ option.

Correct output option

Yes

See below

Broken lead
between
transmitter and
plant data system
or faulty 4-20mA
supply

Verify 4-20mA output
on transmitter and at
data system
Select ‘4-20mA TEST’
under the
DIAGNOSTIC menu to
output discrete 4 to 20
mA signal in 1mA
increments.

Yes, output from
transmitter and input to
plant system

Wiring is good

No signal at plant data
system

Suspect wiring problem

No signal from
transmitter

Suspect faulty 4-20mA output. Acquire
Snapshot, Upload to secure Website, and
contact Customer Support.