2 transmitter mounting instructions, 1 bulkhead mounting, Transmitter mounting instructions -6 – CiDRA SONARtrac PW VF-100 User Manual

Page 72: Bulkhead mounting -6, Figure 1, Bulkhead mount hole pattern -6

Advertising
background image

20832-01 Rev 05

Page 6-6

6.2

Transmitter Mounting Instructions

The transmitter is furnished with a Bulkhead (wall or panel surface
mount) Installation Kit. An optional Pipe Installation Kit is also
available.

The maximum sensor head to transmitter cable length is 300 feet.

Select an installation location that allows for easy and safe access to
the transmitter. Ensure the local ambient temperature range is within
the operating temperature limits of the transmitter. Avoid locations
with extreme vibration and locations that are subject to extreme water
conditions (i.e. direct hose-down). Consider access to power and the
related installation requirements of paragraph 6.3.5, Transmitter
Electrical Cable Installation.

6.2.1

Bulkhead Mounting

The transmitter is attached to the bulkhead or panel with user supplied
1/4-inch fasteners through the four panel mounting feet on the
transmitter. The mounting dimensions are illustrated in the following
figure.

Figure 1

Bulkhead Mount Hole Pattern

Advertising