2 gvf system initial setup, Gvf system initial setup -11 – CiDRA SONARtrac PW VF-100 User Manual

Page 185

Advertising
background image

20835-01 Rev 04

Page 9-11

9.4.2

GVF System Initial Setup

In order to conduct measurements, the ‘Basic Config’ setup must be
completed. If power goes off, this setup will remain in memory and
does not have to be re-entered.

Press any key except ‘EXIT’ on the keypad to enter the ‘Menu
Mode

’.

When ‘Basic Config’ menu is displayed on Line 1 of the display
press the ‘ENTER’ key to enter the options available under that
menu.

→Sensor Serial #’ will appear on Line 2 of the display. Press the
ENTER’ key and the current sensor serial number entered in the
transmitter will be displayed on Line 4 of the display. The user can
then use the arrow keys (

and to scroll through the values and

and to change position) to enter the sensor band serial
number found on the sticker attached to the sensor band. Once all
digits are entered, press the

ENTER’ key to save to memory.

→Sensor Serial #’ will again be displayed on Line 2.

Next, press the

key to scroll to the ‘Pipe Size’ menu on Line 2.

Press the

ENTER’ key to access the options available under that

menu. Note: it is necessary to select only one of the following
options.

The first choice on the

Pipe Size’ menu is ‘→ID / Wall’ shown on

Line 3 of the display. If this value is known it can be entered here.
Press the

ENTER’ key and the current saved inner diameter and

wall thickness will be shown on Line 4 of the display. Use the
arrow keys (

and to scroll through the values and and to

change character) to enter the pipe inner diameter. Units can be
inches or millimeters. Once all digits are entered, press the
ENTER’ key to save the value to memory.

The second choice on the

Pipe Size’ menu is ‘Size/Sched’ (pipe

Size / Schedule). If this value is known, press the

ENTER’ key.

Use the arrow keys to enter the values and then press the
ENTER’ key to save to memory.

The third choice on the

Pipe Size’ menu is ‘OD / Wall’ displayed

on Line 3. Use the arrow keys (

and to scroll through the values

and

and to change character) to enter the pipe inner

diameter. Units can be inches or millimeters. Once all digits are
entered, press the

ENTER’ key to save the value to memory.

Following the entering of ‘Pipe Size’ press the ‘BACK’ key and
→Pipe Size’ will be displayed on Line 2. At this point the user can
either press the

ENTER’ key to re-enter the ‘Pipe Size’ menu or

press the

key to move to the next menu item.