System faults (page 2) – CiDRA SONARtrac PW VF-100 User Manual

Page 262

Advertising
background image

20838-01 Rev 09

Page J-3

System Faults (page 2):

Item
No

PROBLEM

POSSIBLE
CAUSE

DIAGNOSTIC
METHOD

DIAGNOSTIC
RESULT

COMMENTS

3

The display
has random
lines or
makes no
sense

BOOT and / or
MAIN software is
corrupt

Display shows lines or
random characters

---

Contact Customer Support.

4

System
keeps
turning on
and off

Low or poor
quality power to
system

Verify power source for
transmitter is stable
and within operating
range of system

Power fluctuates

Connect to stable power source

Power is good and
stable at between 100

240 volt (AC units) or 18
– 36 volt (DC units)

See below

Poor connections
at ‘Power Input’
terminal block in
transmitter

Verify power
connections to
transmitter

Wire(s) not properly
installed

Reconnect power cable

Wires installed properly

Problem may be internal to transmitter.
Contact Technical Support.

5

The system
keeps re-
booting on its
own

Poor power
source for system

Verify power source for
transmitter is stable
and acceptable

Power fluctuates

Connect to stable power source

Power is good and
stable at between 100

240 volt (AC units) or 18
– 36 volt (DC units)

See below

System memory
error or problem

Run ‘RAM TEST’ and
‘DPRAM TEST’ in
‘SELF TEST’ in
DIAGNOSTICS menu

Fail either or both RAM
and DPRAM test

Faulty RAM/DPRAM.
Contact Customer Support.

Pass both RAM and
DPRAM tests

Acquire Snapshot, Upload to secure Website,
and contact Customer Support.