9.1

Initial Start-up

At initial power up the Green LED will light and the transmitter will
display startup progress; load firmware into the system, load system
parameters, and the heartbeat indicator will activate as the transmitter
begins to collect data from the sensor head. Once startup is
complete, the display will clear and begin displaying measurements.
The following figure illustrates an initial system startup screen.

Figure 1

System Start Display

The display will clear and show the ‘Startup >>>>’ if internal
configuration parameters are changed in the menu screens. This
indicates that the system is restarting with the new parameters and is
beginning to make measurements. Once the startup procedure is
complete, the system measurements will be displayed.

WARNING

When explosive gases may be present, the transmitter door may
be opened only to use the keypad or reset button. Obtain hot
work permit and ensure explosive gasses are not present prior
to performing any other operation.

The SONAR PROCESS MONITORING SYSTEM SUPPLEMENT
FOR ATEX ZONE 2 SAFETY provides additional information for
ATEX Class I, Zone 2 installations.

Startup >>>>>>>>>>>>>>>>

010.001.009.1
34

System is
starting

Firmware
revisions as
they load:
I: Interface
F: FPGA
D: DSP
P: PreAmp

IP Address

Mode and
heartbeat

D 01.00.23 P

/