20838-01 Rev 09

Page A-4

Sensor to Transmitter Cable

The standard sensor to transmitter cable consists of 12 twisted pairs
of 20 AWG conductors with an overall shield, encased in a PVC
jacket. The standard cable has an operating range of -4ºF

– +221ºF

(

–20ºC to +105ºC). The cable is UL Listed (UL Standard 13, Type

PLTC) and CSA Certified (CSA C22.2 No. 214, PCC FT4). The outer
diameter of the cable is 0.61 inch (15.5 mm) nominal.

Optional low temperature and armored cables are also available.
Please contact Customer Support for more information.

Cable Type

25 ft cable weight

Weight per additional foot

lbs

kg

lbs

kg

Standard

5.75

2.61

.23

.10

Armored

12.6

5.72

.45

.20

Sensor to Transmitter Cable Weights

The passive sonar flow meter is configured such that the transmitter is
always located remotely from the sensor head. The cable connecting
the sensor head to the transmitter can be up to 300 feet long.

Electrical Connections

Holes sized for 3/4 inch NPT (M25) fittings (1-1/16 inch diameter) are
located on the base of the transmitter enclosure. The sensor to
transmitter cable is terminated to the screw terminal block within
transmitter enclosure.

Note: All connector glands used on the transmitter enclosure should
be rated to NEMA 4X to maintain transmitter rating of NEMA 4X. For
ATEX Zone 2 Transmitters, further gland requirements can be found
in SONAR PROCESS MONITORING SYSTEM SUPPLEMENT FOR
ATEX ZONE 2 SAFETY.

A single cable connection is made between the transmitter and the
sensor head. Sensor head connection is made with a connector that
is factory installed on one end of the supplied cable. There are two
different sizes of connector used on passive sonar sensor heads. The
cable supplied with the system will have the properly sized mating
connector.