Flow meter faults (page 3) – CiDRA SONARtrac PW VF-100 User Manual

Page 269

Advertising
background image

20838-01 Rev 09

Page J-10

Flow Meter Faults (page 3)

Item
No

PROBLEM

POSSIBLE
CAUSE

DIAGNOSTIC
METHOD

DIAGNOSTIC
RESULT

COMMENTS

14

Flow does
not seem
correct

A cable connection
between the sensor
and transmitter is cut
or disconnected

Run ‘SENSOR CHECK’
under DIAGNOSTIC
menu

Pass

This indicates all sensors are good and
functional.

Fail

1. Verify wire connections in transmitter are
all good.
2. Verify cable is not cut or damaged.
3. Verify cable connector on sensor head is
connected.
4. Remove sensor head access panel per
Installation Manual and ensure sensor to pre-
amplifier connector is properly connected.
5. If system does not function, acquire
Snapshot, upload to secure Website, and
contact Customer Support.

15

The flow rate
displayed is
a series of
dashes ( -----
- )

System ‘Quality
Metric’ is below set
point

The default value of VQ
(if displayed on the
transmitter front panel)
is 0.2.

Value is equal to or
greater than 0.2
consistently

Indicates system is above minimum threshold
for making a reading

Value is less than 0.2

Indicates system is below minimum threshold
to make a reading.
Acquire Snapshot, Upload to secure Website,
and contact Customer Support.

Sensors are
saturating

Go to ‘EVENT LOG’ in
INFO menu and press
ENTER

Yes - A message of
‘Sensor Over’ will be
present if a sensor has
been saturated

1.

Go to ‘GAIN’ on DIAGNOSTICS menu.

2. Press ENTER and scroll to AUTOSET
GAIN and press ENTER
3. GAIN will be adjusted to proper level

No message for
‘Sensor Over’

See below