5.7.1.1

Cover Installation Procedure

Care must be taken during installation of the sensor band cover to
ensure the sensor band cable does not become pinched between the
cover halves. The problem may show up as a sensor failure during
sensor tests and operation of the meter. This potential problem is
most likely to occur in small size meters (<6-inch) due to the length
and stiffness of the sensor band cable.

5.7.1.2

Horizontal Pipe Installation

Sensor cover installation on horizontal pipes is as follows:

1. Install the sensor band.

2. Remove the sensor cable access panel from the sensor upper

cover assembly.

3. Install the sensor upper cover assembly over the sensor band

assembly.

4. Install the sensor lower cover assembly.

a. Hold the halves in place using the cover latches and strikes as

the cover halves are aligned. Note: Ensure the sensor cable
connector is accessible through the sensor cable access panel.
(If necessary, reposition cover or cable connector.)

5. Align the cover halves.

6. Start at the center of the cover and engage the latches and

keepers using a ¾-inch wrench or socket to lock them together
until all are attached.

Figure 20 Fiberglass Cover Assembly

7. Install sensor band boot seal clamps per Section 5.7.1.2.

Cover latches
(typ 6 – 8 plcs)

Boot seal

Lifting
handle

IP-65 rated
connector

Electronics
access panel

Breather
valve

Upper cover
assembly

Lower cover
assembly