System faults (page 5) – CiDRA SONARtrac PW VF-100 User Manual

Page 265

Advertising
background image

20838-01 Rev 09

Page J-6

System Faults (page 5):

Item
No

PROBLEM

POSSIBLE
CAUSE

DIAGNOSTIC
METHOD

DIAGNOSTIC
RESULT

COMMENTS

9

The Display
is Frozen
(reads same
value all the
time)

Transmitter left in
Idle Mode

Is Heart Beat Indicator
Turning in Lower Right
Corner

Yes

– Heart Beat

Indicator is Turning

Acquire Snapshot, Upload to secure Website,
and contact Customer Support.

No- Heart Beat Indicator
is not turning

Cycle power to transmitter

Transmitter is
hung up

Cycle power to
transmitter. Does the
display update?

Yes

– Display updates

Meter is properly functioning

No

– Display is still

frozen

Acquire Snapshot, Upload to secure Website,
and contact Customer Support.

10

The Display
Reads
Preamp
Failure and
there is no
flow reading

Possible faulty
Preamp or bad
connection

Verify wiring and
connectors are
properly connected

No

– wiring / connector

fault found

Correct fault

Yes

– all wiring and

connectors is correct

Acquire Snapshot, Upload to secure Website,
and contact Customer Support.