CiDRA SONARtrac PW VF-100 User Manual

Page 133

Advertising
background image

20833-01 Rev 04

Page 7-39

Table 11 (page 7) GVF Transmitter Menu Tree Software Release 04.02.XX

Level 1

Level 2

Level 3

Range

Description

Diagnostics

Sensor
Check

PASS or FAIL.
(Indicate which sensors
failed test)

Performs health check
on each sensor

4-20mA Test

Test 4-20mA outputs
from 4 to 20mA

Manual testing of 4-
20mA output #1 and 2

Gain

Autoset Gain

1.0, 4.65, 21.55, 98.65

Provides auto
adjustment of pre-
amplifier setting based
on the current process
operating condition

Check/Set
Gain

1.0, 4.65, 21.55, 98.65

Provides for manual
checking and setting of
pre-amplifier gain.
Autoset Gain can
override set point

Test Gain

PASS or FAIL

Test preamp to
determine if gain falls
within range of AGC
parameters

Self Test

RAM Test

PASS or FAIL

System memory test

DPRAM Test

PASS or FAIL

Dual port memory test

Keyboard
Test

Red LED illuminates
indicating key is
functional

Test keyboard operation

Clear History

Reset

Resets data history

Monitor

Passkey

Factory Technical
Support diagnostic
feature

System
Monitor

Sensor
Monitor

Pulse Test

Provides for testing the
pulse output

Alarm Test

Provides for testing the
alarm output

Advertising