Figure 12 – CiDRA SONARtrac PW VF-100 User Manual

Page 35

Advertising
background image

20831-01 Rev 05

Page 5-17

Figure 12 Sensor Band Spacing Tool Installed on Sensor Screw

Note: Ensure the spacer tool is perpendicular to the attachment rail to
ensure proper tightness. Remove the tool, move to the next sensor
screw, and repeat the tightening on each of the sensor screws.

Important: Tighten each screw once only. Do not retighten each
screw using the gauge.

Final sensor band screw tightening is as follows:

A. For sensor bands sized for 6” and smaller pipe:

1. Tighten screws #1-7 an additional one-half turn in the numbered

sequence given in Figure 9. Do not tighten screw #8 & 9 (screws
on either end of the sensor band).

B. For sensor bands sized for 8” and larger pipe:

1. Starting at screw #1 in Figure 9, tighten each screw an additional

one-half turn in the given numbered sequence.

2. Once all nine screws have been tightened, tighten each screw an

additional one-half turn in the given numbered sequence.

3. Once all nine screws have been tightened a second time, tighten

screws #1-7 an additional one-half turn in the given numbered
sequence.

Attach the sensor cable in the retaining clip on the top of the sensor
band. The connector on this cable will attach to the sensor cover
connector as described later in the manual. If necessary, use a piece
of tape to temporarily retain the connector on the band so it will be
positioned directly below the sensor cable access panel. The final
sensor band assembly is illustrated below.

Sensor screw spacing tool
installed on sensor screw

Sensor screw with
Belleville washers
compressed (typ)

Side view