Sensor Assembly Cable Connection

Remove the tape (if it was used) that was temporarily installed to
retain the sensor connector under the cover access panel. Remove
the factory installed shorting plug (if so equipped) from the sensor
cable connector. Save this shorting plug as it will be used if sensor
removal from the pipe is necessary. [Note: Sensor bands with model
numbers ending in “-R” do not require shorting plugs and are delivered
without them.] Install the sensor assembly cable connector into the
mating receptacle located inside the access cover plate on the sensor
upper cover as shown below. The connector is installed such that the
connector key faces away from the pipe. Push the connector into the
mating receptacle. The latches on the receptacle will engage to keep
the plug from pulling out. Ensure the connector is fully engaged.
Note: Use care to ensure the sensor cable connector goes on
squarely to minimize risk of damaging the contacts.

Figure 45 Sensor Assembly Cable Installation

Re-install the sensor cable access panel on the sensor top cover
assembly. Tighten the six panel screws to a recommended torque of
11-lb

f

-in (1.2-Nm).

Connector latches

Sensor cable
access port

Sensor cable with key
positioned up away from pipe