20832-01 Rev 05

Page 6-23

to the connector body and then final tighten 1-1/2 rotations using 1-
5/8” wrenches. Hand-tighten and then final tighten the
compression nut to the connector body 1 rotation using 1-

5/8”

wrenches. Cut and remove the conductor outer sheath about 3/4"
(19mm) from the end of the entry component. Remove foil outer
wrap and foil from each pair of conductors. As each pair is
unwrapped, twist each pair of conductors to keep them together as
pairs.

Figure 18

Armored Cable Connector Installed

Install the seal washer on the connector assembly entry
component. Insert the cable and entry component into the middle
hole in the transmitter box (stiffener plate previously installed.)
Install the ground ring wire assembly and lock nut. Secure the
connector assembly with the lock nut. Attach the ground wire from
the connector nut to any available SHD terminal on the Section #3
terminal block. Strip and install the individual connectors and drain
wire per non-armored cable installation instructions.

Figure 19

Armored Cable Installation

12 twisted pairs
plus drain wire

Connector
body

Compression
nut

Armored
cable

Seal washer

Ground ring wire

assembly

Lock nut

Entry
component

Connector body

Stiffener plate

Ground ring wire assembly

Conductor wires

Seal washer

Lock nut