520 Indicator Installation Manual

2.4.1

Load Cells

To attach cable from a load cell or junction box, route

the cable to the external J6 connector. Wire the load

cell cable from the load cell or junction box to

connector J6 as shown in Table 2-2. If using 6-wire

load cell cable (with sense wires), open the enclosure

cover (see Section 2.3) and remove jumpers JP1 and

JP2 before reinstalling connector J6. For 4-wire

installation, leave jumpers JP1 and JP2 on.

Table 2-2. J6 Pin Assignments

Connector

Pin

Function

J6

1

+SIG

2

–SIG

3

+SENSE

4

–SENSE

5

+EXC

6

–EXC

7

SHIELD

For 6-wire load cell connections, remove
jumpers JP1 and JP2.

2.4.2

Panel Mount Parts Kit, PN 82427

Power connections to the 520 Panel Mount are shown

below. This only applies to the panel mount.
Part number 15888 is used to connect AC power to

balloon number 12 of Figure 2-9 on page 15.
Attach the wires per the diagram shown below.

RICE LAKE WEIGHING SYSTEMS

U1 9

F2

J4

J5

F1

J

1

2

3

4

5

6

7

1

2

3

4

5

6

7

8

9

10

Hot

AC Gr

ound

Neutral

Figure 2-5. AC Power Cord Connection Location

2.4.3

Serial Communications

Attach serial communications cables to external

connector J5. Connector J5 provides connections for

the EDP (Electronic Data Processing) port, printer port,

20 mA current loop transmit signals, and RS-485

signals. Table 2-4 shows the pin assignments for

connector J5.

2.4.4

Condec UMC2000 Conversion Connections

See Table 2-3 for information on load cell connections

when replacing the Condec UMC2000 indicator with

the 520 desktop model indicator.

Note

Pin-outs and connectors for serial
communications and BCD are the same for
both units.