12.12 Analog Output Calibration

The following calibration procedure requires a multi

meter to measure voltage or current output from the

analog output module. If the option is not already

installed, see Section 2.4 on page 8.

Note

The analog output must be calibrated after the
indicator itself has been configured
(Section 3.0) and calibrated (Section 4.0).

9999999

9999999

"-(065

%*(*/

4&5154

9999999

130(3.

1'03.5

4&3*"-

$"-*#3

9999999

$0/'*(

'03."5

4063$&

(3044





0''4&5

&33"$5

'6--4$

)0-%



OVNCFS

.*/



OVNCFS

."9

58;&30

/&5

5841"/

;&304$

9999999

7&34



OVNCFS



OVNCFS

.*/ /&(

."9/&(

0/

0''

0/

0''

Figure 12-4. Analog Output Menu

1. Enter setup mode and go to the ALGOUT

menu (see Figure 12-4):

Set OFFSET to 0% for 0–10 V output,

20% for 4–20 mA output

Set mode select jumper on analog output

board to “V” for 0-10 V output, or “I” for

4-20 mA output

Set MIN to lowest weight value to be

tracked by the analog output

Set MAX to highest weight value to be

tracked by the analog output

2. Connect multimeter to connector J1 on the

analog output board:

For voltage output, connect voltmeter

leads to pins 3 and 4

For current output, connect ammeter leads

to pins 1 and 2

3. Adjust zero calibration: Scroll to the

TWZERO parameter. Press to view zero

value, then check voltage or current reading on

multimedia. Press and hold or to adjust

the zero value up or down.

4. Adjust span calibration: Scroll to the TWSPAN

parameter. Press to view span value, then

c h e c k v o l t a g e o r c u r r e n t r e a d i n g o n

multimeter. Press and hold or to adjust the

span value up or down.

5. Final zero calibration: Return to the TWZERO

parameter and verify that the zero calibration

has not drifted. Press and hold or to

re-adjust the zero value as required.

6. Return to normal mode. Analog output

function can be verified using test weights.

12.13 Test Mode

In addition to normal and setup modes, test mode

provides a number of diagnostic functions for the

520

,

including:

Display raw A/D count

Display digital input states

Reset configuration parameters to default

values

Clear non-volatile (battery backed) storage

Set analog output state to zero or full scale

Set A/D offset and gain calibration

Print configuration

To enter test mode, press and hold the setup switch

until the front panel display shows the word

TEST

. After

a b o u t t h r e e s e c o n d s , t h e t e s t m o d e d i s p l a y

automatically shifts to the first test menu function,

A/DTST.

Figure 12-5 on page 87 shows the Test Menu structure;

Figure 12-6 shows the front panel key functions in test

mode. Note that, because the Test Menu functions are

all on a single menu level, the

GROSS/NET

( ) key has

no function. Press the

ZERO

( ) key to exit test mode.

Table 12-12 on page 87 summarizes the test menu

functions.

Important

A/D calibration functions, ADOFFS and
ADGAIN, must be used only by qualified
service personnel, and only after
replacing A/D converter components.
Improper A/D calibration may render the
indicator unusable.