520 Indicator Installation Manual

PROGRM Menu

Parameter

Choices

Description

Level 2 submenus

DATE

DATEFMT
DATESEP
DATESET

Allows selection of date format and date separator character. See Level 3 submenu parameter
descriptions.

Use the DATESET parameter or the SD EDP command to set the date. See Section 9.0 on
page 50
for information about using the EDP commands.

TIME

TIMEFMT
TIMESEP
TIMESET

Allows selection of time format and separator character. See Level 3 submenu parameter
descriptions.

Use the TIMESET parameter or the ST EDP command to set the date. See Section 9.0 on
page 50
for information about using the EDP commands.

CONSNUM

0000000
1-9999999

Consecutive numbering. Allows sequential numbering for print operations. The consecutive
number value is incremented following each print operation that includes <CN> in the ticket
format. When the consecutive number is reset, it is reset to the value specified on the
CONSTUP parameter, up to seven characters.

CONSTUP

0000000
1-9999999

Specifies the consecutive number start-up value used when the consecutive number is reset
by sending the KCLRCN EDP command or a CLRCN digital input. Value specified must be in
the range 0–9 999 999.

UID

0
unit ID

Specifies the unit identification number. Value specified can be any alphanumeric value, up to
seven characters.

ACCUM

ON
OFF

Accumulator. Specifies whether the accumulator is enabled. If enabled, accumulation occurs
whenever a print operation is performed.

RTZ GRADS

0

Return to zero grads allows a tolerance in grads in which the scale can Return To Zero+RTZ
Grads to trigger the next accumulation.

CKWEIGH

OFF
ABS
PERCENT

Checkweigh. Specifies whether checkweigh is enabled and whether set for absolute or
percent checkweigh. If set to a value other than OFF, setpoints are disabled. Digital output one
will be active (ON) when the weight is within the UNDER range, digital output two will be active
when the weight is in the ACCEPT range, and digital output three is active when the weight is
within the OVER range (see Section 7.0 on page 47).

COUNT

ACCESS
SPLSIZ

Specifies whether operator has access to piece count mode and the default sample size used
for parts counting (see Section 6.0 on page 46). See Level 3 submenu parameter descriptions.

PK HOLD

OFF
NORMAL
BI-DIR
AUTO

Peak hold function. Optional feature used to hold display of the highest net weight achieved
during a weighing cycle (see Section 8.0 on page 49).

GRAVADJ

OFF
ON

Gravity adjustment. Set LATITUD and ELEVATN parameters for the geographic location where
the calibration is to take place (see Figure 3-7 on page 25 for these parameters).
Compensates for the variance in gravitational pull from one location to another.

MVCALIB

OFF
ON

Millivolt calibration. Allows entry of calibration values in mV.

CFGPWD

0
1–9999999

Configuration password. Specify a non-zero value to restrict access to all configuration menus.

SPPWD

0
1–9999999

Setpoint password. Specify a non-zero value to restrict access to the setpoint and checkweigh
menus in operate mode.

KYBDLK

OFF
ON

Keyboard lock. Specify ON to disable all front panel keys except SETUP.

ZERONLY

OFF
ON

Zero key only. Specify ON to disable all front panel keys except ZERO in normal mode.

NOTE: Only the ZERONLY parameter can be specified as ON. Individual keys can be locked or
unlocked using the KLOCK and KUNLOCK EDP commands (see Table 9-1 on page 50 for
more information).

Table 3-6. Program Menu Parameters