Calibration

42

4.2

EDP Command Calibration

To calibrate the indicator using EDP commands, the

indicator EDP port must be connected to a terminal or

personal computer. See Section 2.4.3 on page 9 for

EDP port pin assignments; see Section 9.0 on page 50

for more information about using EDP commands.
Once the indicator is connected to the sending device,

do the following:

1. Place the indicator in setup mode (display

reads

CONFIG

) and remove all weight from the

scale platform. If your test weights require

hooks or chains, place the hooks or chains on

the scale for zero calibration.

2. Send the WZERO EDP command to calibrate

zero. The indicator displays

*CAL*

while

calibration is in progress.

3. Place test weights on the scale and use the

WVAL command to enter the test weight value

in the following format:

WVAL=nnnnnn<CR>

4. Send the WSPAN EDP command to calibrate

span. The indicator displays

*CAL*

while

calibration is in progress.

5. Up to five linearization points can be

c a l i b r a t e d b e t w e e n t h e z e r o a n d s p a n

c a l i b r a t i o n v a l u e s . U s e t h e f o l l o w i n g

commands to set and calibrate a single

linearization point:

WLIN.V1=nnnnn<CR>
WLIN.C1<CR>

The WLIN.V1 command sets the test weight

value (nnnnn) for linearization point 1. The

WLIN.C1 command calibrates the point.

Repeat using the WLIN.Vx and WLIN.Cx

commands as required for additional

linearization points.

6. To remove an offset value, clear all weight

from the scale, including hooks or chains used

to hang test weights, then send the REZERO

EDP command. The indicator displays

*CAL*

while the zero and span calibrations are

adjusted.

7. Send the KUPARROW or KEXIT EDP

command to exit setup mode.

4.3

Revolution

®

Calibration

To calibrate the indicator using Revolution, the

indicator EDP port must be connected to a PC running

the Revolution configuration utility.
Use the following procedure to calibrate the indicator:

1. Place the indicator in setup mode (display

reads

CONFIG

) and remove all weight from the

scale platform.

2. Select

Calibrate Indicator

from the Revolution

main menu.

3. On the Indicator Calibration display, select the

indicator model (

520

) and communications

port then click

OK

(see Figure 4-3).

Figure 4-3. Revolution Calibration Display

4. Revolution uploads calibration data from the

indicator then presents the information in a

display.

5. Enter the

Value of Test Weight

to be used for span

calibration then click

START

.

6. A dialog box asks whether hooks or chains are

used to hang calibration test weights. Click

Yes

or

No

to continue.

7. The Zero Calibration dialog box prompts you

to remove all weight from the scale. Clear the

scale and click

OK

to begin zero calibration. If

your test weights require hooks or chains,

place the hooks or chains on the scale for zero

calibration.

8. When zero calibration is complete, the Span

Calibration dialog box prompts you to place

test weights on the scale for span calibration.

Place test weights on the scale then click

OK

.

9. When span calibration is complete, a dialog box

asks whether you wish to perform linear

calibration. Click

Ye s

to perform linear

calibration for up to five linearization points, or

click

No

to continue calibration with step 11.

10. On the Linear Calibration display, select the

point (1–5) to calibrate, then click

Calibrate

.

Place test weights on scale then click

OK

.

When prompted, enter the test weight value

then press Enter. Repeat for additional

linearization points, then click

Exit

to return to

the Indicator Calibration display.