51

520 Indicator Installation Manual

9.1.2

Reporting Commands

Reporting commands (see Table 9-2) send specific

information to the EDP port. These commands can be

used in both setup mode and normal mode.

Table 9-2. EDP Reporting Commands

Command

Function

DUMPALL

List all parameter values

SPDUMP

Print setpoint configuration

VERSION

Write

520

software version

P

Write current displayed weight with units
identifier. See Section 12.4 on page 74 for
more information.

S

Write one frame of stream format

9.1.3

The RESETCONFIGURATION Command

The RESETCONFIGURATION command can be used

in setup mode to restore all configuration parameters to

their default values.
This command is equivalent to using the DEFAULT

function in TEST mode. See Section 12.3 on page 74

for more information about test mode.

Note

All load cell calibration settings are lost when
the RESETCONFIGURATION command is run.

9.1.4

Parameter Setting Commands

Parameter setting commands allow you to display or

change the current value for a particular configuration

parameter (Tables 9-3 through 9-12).
Current configuration parameter settings can be

displayed in either setup mode or normal mode using

the following syntax:

command<ENTER>

Most parameter values can be changed in setup mode

only; setpoint parameters listed in Table 11-2 on

page 69 can be changed when in normal weighing

mode.
Use the following command syntax when changing

parameter values:

command=value<ENTER>

where

value

is either a number or a parameter value.

Use no spaces before or after the equal (=) sign. If you

type an incorrect command, the display reads

??

.

For example, to set the motion band parameter to 5,

type the following:

MOTBAND=5D<ENTER>

For parameters with selectable values, enter the

command and equal sign followed by a question mark:

command=?<ENTER>

to see a list of those values. The indicator must be in

setup mode to use this function.

Table 9-3. CONFIG EDP Commands

Command

Description

Values

GRADS

Graduations

1–9999999

ZTRKBND

Zero track band

OFF, 0.5D, 1D, 3D

ZRANGE

Zero range

1.9%, 100%

MOTBAND

Motion band

1D, 2D, 3D, 5D, 10D, 20D, OFF

OVRLOAD

Overload

FS+2%, FS+1D, FS+9D, FS

SMPRAT

Sample rate

60HZ, 120HZ, 30HZ, 7.5HZ, 15HZ

DIGFLTR1
DIGFLTR2

Digital filtering

1, 2, 4, 8, 16, 32, 64, 128, 256

RATLTRAP

Rattletrap filtering

OFF, ON

DFSENS

Digital filter cutout sensitivity

2OUT, 4OUT, 8OUT, 16OUT, 32OUT, 64OUT, 128OUT

DFTHRH

Digital filter cutout threshold

NONE, 2DD, 5DD, 10DD, 20DD, 50DD, 100DD, 200DD, 250DD

PWRUPMD

Power up mode

GO, DELAY

TAREFN

Tare function

BOTH, NOTARE, PBTARE, KEYED