41

520 Indicator Installation Manual

4.1

Front Panel Calibration

To calibrate the indicator using the front panel, do the

following:

1. Place the indicator in setup mode (display

reads

CONFIG

) and remove all weight from the

scale platform. If your test weights require

hooks or chains, place the hooks or chains on

the scale for zero calibration.

2. Press until the display reads

CALIBR

(see

Figure 4-1). Press to go to zero calibration

(

WZERO

).

3. Standard Calibration: With

WZERO

displayed,

press to calibrate zero. The indicator

displays

*CAL*

while calibration is in progress.

When complete, the A/D count for the zero

calibration is displayed. Press again to save

the zero calibration value and go to the next

prompt (

WVAL

).

Millivolt Calibration: To enable millivolt

calibration, set MVCALIB=ON on the

PROGRM

menu (see Section 3.2.5 on page 30).

Once enabled, press to display and edit

millivolt value at zero weight (see Figure 4-2

on page 41), press again to calibrate and

display raw A/D count, press the third time

to move to the next selection.

4. With

WVAL

displayed, press to display the

stored calibration weight value. Use the

procedure shown in Figure 4-2 to enter the

actual value of the test weights to be used for

the span calibration. Press to store the

entered

WVAL

value and advance to the next

prompt (

WSPAN

).

8IFOFEJUJOHOVNFSJDWBMVFT QSFTT

PS

UPDIBOHFUIF

EJHJUTFMFDUFE1SFTT

PS

UPJODSFNFOUPSEFDSFNFOUUIF

WBMVFPGUIFTFMFDUFEEJHJU*GBEFDJNBMQPJOUJTSFRVJSFEJOWBMVF
QSFTTUPWJFXBOENPWFEFDJNBMQPJOU1SFTT PS UP
NPWFEFDJNBMQPTJUJPO1SFTT UPTBWFUIFWBMVFFOUFSFEBOE
SFUVSOUPUIFMFWFMBCPWF







Figure 4-2. Editing Procedure for Numeric Values

5. With

WSPAN

displayed, place test weights on

the scale and press to calibrate span. The

indicator displays

*CAL*

while calibration is in

progress. When complete, the A/D count for

the span calibration is displayed. Press to

store the span calibration value and advance to

the next prompt (

WLIN

).

Millivolt Calibration: If millivolt calibration is

enabled, press to display and edit millivolt

value at span weight, press again to calibrate

and display raw A/D count, press the third

time to move to the next selection.

6. Five-point linearization (using the WLIN

parameter) provides increased scale accuracy

by calibrating the indicator at up to five

additional points between the zero and span

calibrations.
Linearization is optional: if you choose not to

perform linearization, press the right arrow and

you will keep scrolling through the menu

choices. When complete, use the up arrow to

exit. If linearization values have previously

been entered, these values are reset to zero

during calibration. To perform linearization,

follow the procedure below:

Standard Calibration: With

WLIN

displayed,

Press to go to the first linearization point

(

PT-> 1

). Place test weights on the scale and

press or

ENTER

.

Use the procedure shown in Figure 4-2 to enter

the actual test weight value, then press to

calibrate. The indicator displays

*CAL*

while

calibration is in progress. When complete, the

A/D count for the linear calibration is

displayed. Press again to save the calibration

value and go to the next prompt (

PT-> 2

).

Repeat for up to five linearization points. To

exit the linearization parameters, press to

return to WLIN.
Millivolt Calibration: If millivolt calibration is

enabled press to display and edit the test

weight value, press again to display and edit

the millivolt value for that weight, press

again to calibrate and display the raw A/D

count, press again to move to the next

calibration point.

7. The rezero function is used to remove a

calibration offset when hooks or chains are

used to hang the test weights.

If no other apparatus was used to hang the

test weights during calibration, remove the

test weights and press to return to the

CALIBR menu.

If hooks or chains were used during

calibration, remove these and the test

weights from the scale. With all weight

removed, press to rezero the scale. This

function adjusts the zero and span

calibration values. The indicator displays

*CAL*

while the zero and span calibrations

are adjusted. When complete, the adjusted

A/D count for the zero calibration is

displayed. Press to save the value, then

press to return to the CALIBR menu.

8. Press until the display reads

CONFIG

, then

press so that the display reads

WAIT

, and

returns back to weigh mode.