520 Indicator Installation Manual

Figure 12-5. Test Menu

Move LEFT

Move RIGHT

EXIT

not used

ENTER

ENTER SAMPL

E

MODE

Figure 12-6. Front Panel Key Functions in Test Mode

Table 12-12. Test Menu Functions

TEST Menu

Function

Description

A/DTST

Display A/D test
Press and hold ENTER key to display raw count from A/D converter.

ADOFFS

A/D offset calibration (-0.5 mv/V).

Press and hold the setup switch, then press the ENTER key to perform offset calibration.

ADGAIN

A/D gain calibration (4.5 mv/V).

Press and hold the ENTER key to perform gain calibration.

D/I 1 — D/I 3

Display digital inputs 1–3

Press and hold ENTER key to show status of DIGINx (DIx=HI or DIx=LO).

AOUT=0

Set analog output to zero

Press and hold ENTER key to set analog output to its zero value.

AOUTFS

Set analog output to full scale

Press and hold ENTER key to set analog output to its full scale value.

DEFLT

Default parameters

Press and hold the setup switch, then press the ENTER key to reset configuration and calibration
parameters to factory default values. Load cells must be recalibrated before using the indicator (see
Section 4.0 on page 40);

CLR NV

Clear non-volatile storage

Press and hold the setup switch, then press the ENTER key to clear values stored in battery-backed SRAM.

PRTCFG

Print configuration

Press and hold ENTER key to send configuration data to the serial port. This function is equivalent to using
the DUMPALL EDP command.