Table 3-7. Setpoint Menu Parameters

SETPTS Menu

Parameter

Choices

Description

Level 2 submenus

SETPT1–SETPT8

OFF
GROSSSP
NETSP
+RELSP
–RELSP
%RELSP
PAUSE
DELAY
WAITSS
COUNTER
AUTOJOG
COZ
INMOTON
INRANGE
–GROSS
–NET
BATCHPR
TIMER
CONCUR

Specifies the setpoint kind.

GROSSSP, NETSP, +RELSP, –RELSP, and %RELSP setpoint kinds can be used as
either batch or continuous setpoints.

PAUSE, DELAY, WAITSS, COUNTER, and AUTOJOG setpoint kinds can only be
used in batch sequences.

COZ, INMOTON, INRANGE, –GROSS, –NET, BATCHPR, TIMER, and CONCUR
setpoint kinds can only be used as continuous setpoints.

See Table 11-1 on page 65 for more information about setpoint kinds.

BATCHNG

OFF
AUTO
MANUAL

Batching enable. Set to AUTO or MANUAL to allow a batch sequence to run.
MANUAL requires a BATSTRT digital input, BATSTART EDP command before the
batch sequence can run. AUTO allows batch sequences to repeat continuously.

Note

37

520 Indicator Installation Manual

See Section 11.0 on page 65 for detailed information on setpoints.