Configuration

32

PROMPTS

PRMPT1–
PRMPT8

Specify prompts for use in setpoint names (up to 16 characters). Prompts are referenced by
the NAME parameter under the SETPTS submenus; prompts appear in the secondary display
area during setpoint execution for batch type setpoints.

REGULAT

NTEP
CANADA
NONE
OIML

Regulatory mode. Specifies the regulatory agency having jurisdiction over the scale site.

• OIML, NTEP, and CANADA modes allow a tare to be acquired at any weight greater than

zero. NONE allows tares to be acquired at any weight value.

• OIML, NTEP, and CANADA modes allow a tare to be cleared only if the gross weight is at

no load. NONE allows tares to be cleared at any weight value.

• NTEP and OIML modes allow a new tare to be acquired even if a tare is already present.

In CANADA mode, the previous tare must be cleared before a new tare can be acquired.

• NONE, NTEP and CANADA modes allow the scale to be zeroed in either gross or net

mode as long as the current weight is within the specified ZRANGE. In OIML mode, the
scale must be in gross mode before it can be zeroed; pressing the ZERO key in net mode
clears the tare.

The value specified for this parameter affects the function of the front panel TARE and ZERO
keys. See Section 12.5 on page 75 for a complete description of TARE and ZERO key
functions for each of the regulatory modes.

REGWORD

GROSS
BRUTTO

Regulatory word for GROSS. Specifies GROSS or BRUTTO.

NEG NET

ON
OFF

SETPT

DECIMAL
NODECML

Sets whether the setpoint values use a decimal point or not.

BUS

LFT
IND

Sets the update rate of the Field Bus cards.

Level 3 submenus

DATEFMT

MMDDYY
DDMMYY
YYMMDD

Specifies the format used to display or print the date.

DATESEP

SLASH
DASH
SEMI

Specifies the date separator character.

DATESET

nnnnnn

Enter date specified by the DATEFMT parameter.

TIMEFMT

12HOUR
24HOUR

Specifies the format used to display or print the time.

TIMESEP

COLON
COMMA

Specifies the time separator character.

TIMESET

nnnnnn

Enter time specified by the TIMEFMT parameter.

ABS

UNDER
OVER
CKZBAND
CKPRINT

Select checkpoints by giving absolute minimum and maximum weight values in primary units.

PERCENT

TARGET
UNDER
OVER
CKZBAND
CKPRINT

Select a target weight in primary units and enter the under/over values as a percent of the
target. Specify values as a whole percent. Default is 10%.

ACCESS

DISABLE
ENABLE

Operator access to piece count mode. Specify DISABLE if piece count mode will not be used.
With access disabled, pressing the GROSS/NET (MODE) key toggles between gross and net
modes only.

PROGRM Menu

Parameter

Choices

Description

Table 3-6. Program Menu Parameters (Continued)