RL-TA Series Operation Manual 9

4 Basic Operation

4.1 Power On/Off

Press the On/Off key to turn the power on and off.

The * sign is displayed when the balance is powered by the AC adapter

and is in standby mode.

The balance will start in the mode it was in when last turned off. For example, if you turn off the balance in

the parts counting mode, it will start in that mode when it is next turned on.

(Refer to “Section 5.2: Description of Functions 1” on page 16 for details of the weighing modes.)

4.2 Weighing

Place a sample on the balance and close the windshield door.

A circle will be displayed after the measurement has stabilized.

Read the measurement while the circle is displayed.

The circle will not be displayed or will flicker if the balance is

subject to disturbance such as vibration. Take appropriate

measures to stabilize the balance.

The bar graph indicates the current load in relation to the

balance’s weighing capacity. As the load approaches the

weighing capacity, the bar extends towards the right.

When the weight of an object exceeds the balance’s weighing capacity, [

o

-

Err] is displayed.

If the pan is removed and the zero-point falls below the original zero-point,

[

u - Err] is displayed.

4.3 Zero Adjustment

Press the Zero/Tare key.

Wait until the display becomes stable (the M sign flashes), and set

the display to “0”.

While the zero point is accurately maintained, the [

→0←] sign is

displayed.

4.4 Tare (pan)

Place the tare (pan) on the balance and press the Zero/Tare key.

When the display becomes stable, “0” is displayed.

(Continued on next page.)