RL-TA Series Operation Manual 41

(Example)

To set a lower limit of 97.0000 g and an upper limit of 105.0000 g with respect to a reference weight

of 100.0000 g, enter the limit values as shown below:

Reference Weight

Lower Limit

Upper Limit

Absolute weight

100.0000 g

97.0000 g

105.0000 g

Judgment by absolute values

100.0000 g

97.0000 g

105.0000 g

Judgment by deviation values

100.0000 g

-3.0000 g

5.0000 g