17.4.3 Command format

(1) Tare range (zero-setting) command

C1

C2

ASCII code

Description

Value

Response

T

(SP)

54H

20H

 Set Tare

Range

 Adjust the

zero-point

None

A00: Successful completion
E01: Command error
E04: A tare range cannot be set or

the zero-point cannot be
adjusted (because of a range
violation or a weight error).

(2) Set output control command

C1

C2

ASCII code

Description

Response

O

0

4FH

30H

Stop output

A00: Successful

completion

E01: Command

error

E02: Interval

time error
(only with
OA or OB)

O

1

4FH

31H

Output continuously at all times

O

2

4FH

32H

Output continuously if stable (Stop output if
unstable)

O

3

4FH

33H

Output once when the Print key is pressed
(whether the balance is stable or unstable).

O

4

4FH

34H

Output once when the balance is stable.
Output when a substance is lifted to cause
the display to indicate a value below zero,
and then another substance is put on to
make the balance stable.

O

5

4FH

35H

Output once when the balance is stable.
Stop output when unstable. Output once
when the balance is stabilized again (the
output includes zero) even if it is not
reloaded.

O

6

4FH

36H

Output once when the balance is stable.
Output continuously when unstable. Output
is stopped after a single output when the
balance is stable even if it is not reloaded.

O

7

4FH

37H

Output once when the Print key is pressed
if the balance is stable.

O

8

4FH

38H

Output once immediately

O

9

4FH

39H

Output once after stabilized

O

A

4FH

41H

Interval function
(Output once each time the output time
elapses.)

O

B

4FH

42H

Interval function
(Output once if stable each time the output
time elapses.)

* Commands O0 to O7 control output in the same way as the settings in Function 1.

* Commands O8 and O9 are used to request data from the balance.
* Once executed, O0 to O7, OA, or OB commands is held until the next command is input. However,

output control is reset to the function setting when the balance is turned on again.

* When command OA or OB is input, the balance initiates the interval function, and when the same

command is input again, it terminates the function.

* When either an O8 or O9 command is executed, it returns to the state of “O0”.