38 RL-TA Series Operation Manual

13.5 Judge by Deviation Values

13.5.1 Set two limit values by putting actual samples on the balance Judge by Deviation Values

1.

Start limit value setting.
Press and hold the Target key for a few seconds.

When [

r.SEt] is displayed, release the key.

The current reference value blinks.

2.

Enter a reference weight.

Put a sample that provides a reference weight on
the balance, and press the Function key.

When a reference weight is saved, the value is

displayed. The next item will then be displayed.

3.

Enter a lower limit value.
The display first indicates [

L. SEt], and then the

current lower limit blinks. Put a sample that

provides a lower limit on the balance, and press
the Function key.

The error between the lower limit value and the

reference weight is displayed. Then, you can set

an upper limit.

4.

Enter an upper limit value.

The display first indicates [

H. SEt], and then the

current upper limit blinks. Put a sample that

provides an upper limit on the balance, and press
the Function key.

The error between the upper limit value and the

reference weight is displayed in the same way as

the setting of a lower limit. Then the balance

returns to measurement mode.