18 use printers, 1 setting up the printer, 2 outputting calibration results – Rice Lake TA Series, Rice Lake Weighing Systems User Manual

Page 74: Set function 1, Header output, Measurement data output, Footer output, Use printers, Setting up the printer, Outputting calibration results

Advertising
background image

70 RL-TA Series Operation Manual

18 Use Printers

18.1 Setting up the Printer

(1) Use a dot matrix or thermal RS-232 printer with the balance.

(2) Set proper print functions (print control) with the balance referring to the instruction manual for your

printer.

(3) Make the baud rate and other settings compatible between the balance and the printer.

18.2 Outputting calibration results

18.3 Output of Measurement Data in Compliance with ISO/GLP/GMP

1.

Set Function 1.

Set the ISO/GLP/GMP-compatible item to [

F GLP 1] and the GLP-compatible item for

measurement data to [

F2. od. 1].

2.

Header output.
Press and hold the Print key for a few seconds.
[

HEAd] is displayed and the header is then output.

3.

Measurement data output.

Measurement data can be output at any time during measurements.

4.

Footer output.
Press and hold the Print key for a few seconds after the

measurement is completed.

The display indicates [

Foot] and a footer is output.

(Continued on next page.)

Advertising