2 net addition function, Reset the display to “0, Put samples on the balance, and obtain the weight – Rice Lake TA Series, Rice Lake Weighing Systems User Manual

Page 36: The balance automatically sets tare range, Net addition function

Advertising
background image

32 RL-TA Series Operation Manual

Obtain a total weight without reloading the balance
Press the Zero/Tare key in step 3 without unloading the balance. This will reset the
display to “0”.
Put additional samples on, and press the Target key. Then, the weight is cumulated

.

1.

Put additional samples on when the display indicates “0”.

2.

When [

t - Err] is displayed when you press the Target key, it indicates that you put additional

samples on twice, that you unloaded some samples, or that you pressed the key without adding

samples.

3.

You can add samples when an asterisk (*) is displayed.

4.

The [

J. tA.] display with Function 1 can be used for turning ON/OFF the function to wait for

stabilization upon additions.

12.2 Net Addition Function

The net addition function automatically performs tare range after an addition, so additions can be

performed without having to replace samples.

Function 1 is [

2. SEL 1] and [26 Ad. N 2].

1.

Reset the display to “0”.
Press the Zero/Tare key.

2.

Put samples on the balance, and obtain the weight.

Put samples to weigh on the balance. When the asterisk (*)

sign is displayed, you can put additional samples on.

When

is displayed, press the

Target key

. The

displayed weight is added, and the cumulative

weight is displayed with a [

Σ] sign for several seconds.

3.

The balance automatically sets tare range.

The balance automatically sets the tare range.

4.

Put additional samples on the balance to obtain the weight.

When

is displayed after adding the sample, press

the

Target key

.

The displayed value is added, and the cumulative weight

is displayed with a [

Σ] sign for several seconds.

(Continued on next page.)