4 input commands, 1 procedure for transmission, Input commands – Rice Lake TA Series, Rice Lake Weighing Systems User Manual

Page 66

Advertising
background image

62 RL-TA Series Operation Manual

Time output

When the time stamp function is enabled, the time is output one line above the output data.

1

2

3

4

5

6

7

8

9

10

h

h

:

m

m

:

s

s

CR

LF

* hh: hour (00-23), mm: minute (00-59), ss: second (00-59)

17.4 Input Commands

The following eight input commands are supported:

(1) Tare range command

(5)

Request time output command

(2) Set output control command

(6)

Set interval command

(3) Set measurement mode command

(7)

Span adjustment/test command

(4) Request date output command

(8)

Setting limit values

17.4.1 Procedure for transmission

(1) An input command is sent from an external device.

The full-duplex transmission system allows you to send an input command at any time independently

of the data transmit timing of the balance.

(2) Upon successful completion of an input command, the balance will send out either a normal

completion response or the data requested by the command. If the operation has not resulted in

normal completion, or if the received input command is invalid (in error), the balance will transmit an

error response.

When the balance is in normal display mode, it usually sends a response to an input command within

one second after it is received.

However, if the balance receives a tare range command when the Set Tare function is set so that the

display is reset to “0” when the balance is stabilized after the Zero/Tare key is pressed, or if it receives

an input command that takes a long time, it sends a response after the command is completely

processed.

If the balance receives an input command when you are setting a function, when it is under span

adjustment, or it is busy for other reasons, it sends a response after that operation is completed.

1) Commands to which a response is sent back after a specified operation is executed:

 Tare range command, span adjustment/test command

2) Commands to which a response is sent back immediately or in one second after they are received:

 Commands other than those specified in (1) above

3) Once you have sent an input command from an external device, do not send another input command

to the balance until the external device receives a response to the command from the balance.

Advertising