8 Parts Counting

When a specified number of samples are put on the balance, the balance divides the total weight of the samples
by the number of samples to obtain the average sample weight. The balance stores the average sample weight
based on the number of samples initially set. Additional samples, up to three times the initial number, can be
added and the balance will automatically update the average sample weight. This mechanism allows accurate
counting.

* Select

[

1. Set 2] (Parts Counting) in Function 1.

1.

Start sampling.

Make sure that the [Pcs] sign is displayed and press and hold
the Function key for a few seconds.

(This operation may be performed with a container placed on

the balance.)

When the display indicates [

U. SEt],

release the key.

The display then blinks [

on 10].

This indicates to put 10 samples on the balance.

2.

Change the number of samples as required.

The number of samples can be changed by pressing the

Zero/Tare key.

If the samples vary considerably in size or are lightweight, or if
more accurate counting is required, press the Zero/Tare key to

sufficiently increase the number of samples.

If you do not need to change the number, skip this step.

3.

Put a sample on the balance.

Put as many samples as displayed in [

on **] on the balance

and press the Function key.

The display then blinks [

10

Pcs

].

4.

Put additional samples on the balance.

Add up to three times the displayed number of samples and

wait until the display stabilizes again. The [+] sign is

displayed and the memory is updated.

Every time the [+] sign is displayed, you may add more samples.