7 data output, 8 hanging measurement, Data output – Rice Lake TA Series, Rice Lake Weighing Systems User Manual

Page 15: Hanging measurement

Advertising
background image

RL-TA Series Operation Manual 11

Target key

Meanings

Description

Corresponding settings in

Function 1

6. EnU.

5. rE.

FAST

The balance responds quickly but it

is easily affected by vibration

0

1

NORMAL

The balance’s response is

somewhere between fast and slow

0

3

SLOW

The balance responds slowly but is

not easily affected by vibration

1

3

(Refer to “Section 5: Function 1” on page 15 for further information on Function 1.)

In

,

[

0] indicates a fast response. Use this setting when the

balance is subject to a low level of vibration.
[

1] indicates a slow response. Use this setting when the balance

is subject to a high level of vibration.

In [

,

[

1] indicates the fastest response. Use this setting when

the balance is subject to a low level of vibration. As the number
approaches [

5], the balance’s response becomes slower but it is

also less affected by vibration.

* To adjust the response more finely, use Function 1.

This function is unavailable when the gravimeter or addition function is enabled.

4.7 Data Output

Measured data and GLP data can be output to a personal computer, printer, or similar device via the
built-in RS-232C interface. In the factory default setting, pressing the Print key once will output the

current measured value when the measurement becomes stable.

is displayed while the data is being output.

Refer to “Section 5.5: Interface” on page 20 for further information on the output settings.

4.8 Hanging Measurement

The balance is equipped with a hanging hook to suspend a hanging pan for weighing. Perform hanging

measurements when weighing electromagnetic, electrostatic, or high-temperature samples that cannot be

accurately weighed on the normal pan.

Remove the pan and gently tilt the balance backward.

Take care that the door does not fall off.

(Continued on next page.)