52 RL-TA Series Operation Manual

16.4 Set Unit

This is a function to allow you to select two units (unit A and unit B) and switch between the units.

Which unit to display can be set in Function 1. You can also set the same unit for unit A and unit B.
Pressing the Function key switches the unit between unit A and unit B.

To use unit B, select the weighing machine function in Function 1 [

1. Set 1]. Unit A can be used in all

the measurement types of the balance.

16.5 Minimum Readability Setting

This function changes the minimum readability. The increment of the minimum readability changes in

the order of 1, 2, and 5. The larger the minimum readability becomes, the less the balance is affected by

disturbance. In addition, it takes less time for the balance reading to become stable.

You can also set the same unit for unit A and unit B in “Section 16.3, Auto Power Off” and switch the

minimum readability between unit A and unit B.

For example, when grams are used the minimum readability is as follows:

Function 1

C2. d.A

Minimum

Readability

C4. d.b

Setting

values

1

0.0001 g

2

0.0002 g

3

0.0005 g

4

0.001 g

5

0.002 g

16.6 Date Display

This is a function to set the date format to be displayed in the balance or output to a printer.

Set date format in Function 1 [

F. dAtE *]:

16.7 Time Stamp Output

This is a function to output measurement data with the current time. You should set the time before this

function is used.

To use this function, enable it in Function 1 [

G. t.o. 1].