12 Addition Function

The addition function adds up the weights of samples put on the balance, allowing you to find the cumulative

weight. Except for the gravimeter function, all other functions, i.e., weighing machine, parts counting,

percentage weighing, and unit converting functions support the addition function.

12.1 Cumulate Function

With the cumulate function, a cumulative weight can be obtained while samples are reloaded.

Select [

2. SEL 1 or 3] and [26. Ad. N 1] in Function 1.

1.

Reset the display to 0.
Press the Zero/Tare key.

2.

Put samples on the balance, and obtain the weight.

Put samples to weigh on the balance. When the asterisk (*)

sign is displayed, you can put additional samples on.

When

is displayed, press the Target key. The displayed

weight is added, and the cumulative weight is displayed with a
[Σ] sign for several seconds.

3.

Put another sample on the balance.

Unload the balance, check that the display indicates “0”, and

put different samples on.

4.

Read the weight.

When

is displayed, press the Target key.

After the cumulative weight is displayed with a [

Σ] sign,

the weight of the sample on the balance will be displayed.

5.

Display the cumulative weight.
Press the Function key several times.

The cumulative weight is displayed with a [

Σ] sign.

6.

Clear the cumulative weight.
Press the Zero/Tare key with the cumulative weight displayed.

Pressing the Function key returns the display to the weight

of the sample on the balance.

(Continued on next page.)