RL-TA Series Operation Manual 47

14.6 Entry of Weight Error

By entering the weight error of an external reference weight to be used in span adjustment or span test, a

more accurate calibration can be performed. Enter the weight error obtained from the following equation:

(Weight errors are entered in the unit of mg.)

Weight Error = Actual Weight - Nominal Value

Example: For a 100 g weight, a weight error may be = 100.00012 - 100

= 0.00012 = +0.12 mg

1.

Select [

2. o.NP. 1] in Function 2.

2.

Display the weight error.

When you only want to check the current weight
error, press the Function key.

[

3. r.CA. 0] for the next item.

3.

Start entering the weight error.
Press the Zero/Tare key.

The leftmost digit blinks first.

4.

Enter a numerical values.
Pressing the Zero/Tare key allows you to change the

blinking value.


Pressing the Function key makes a higher-order digit blink.

After a number is set, the minus sign [M] blinks. Pressing

the key again makes the digit furthest to the right blink.
Pressing the Print key returns the display to Step 2.

5.

Save the weight error
Press the Target key.

([

3. r.CA.0] shown on the next page is displayed [refer to page 47].)

Press the Target key to return the display to measurement mode.

Using a saved weight error

For using the weight error entered, set to Function 2 to [

4. N.E.H. 1].