9 Percentage Weighing

The weight of a reference sample is saved as 100%. When a sample is placed on the balance, the balance

indicates the percentage of that sample with respect to the reference sample weight.

A reference sample weight can be set by weighing an actual sample or entering a value.

* Select [

1. Set. 3] (Percentage Weighing) in Function 1.

9.1 Set a reference weight by weighing an actual sample

You can make the display indicate the weight percentage (%) of a sample with respect to a reference

sample weight.

1.

Start sampling.

Make sure that the [%] sign is displayed and press and hold the

Function key for a few seconds.

(This operation may be performed with a container placed on

the balance.)

The display then blinks [

P. SEt].

The previously-saved reference sample weight blinks.

2.

Put a sample on the balance.

Put the reference sample on the balance.

3.

Save the reference weight.
Press the Function key.

4.

Put a sample to weigh on the balance.

The display indicates the percentage of the sample weight with

respect to the saved reference sample weight.

The minimum unit is automatically selected, depending on the saved reference weight.

Min. Indication

Range of Reference Weight

L -- Err

Reference weight < Lower weight limit

1%

Lower weight limit

≤ Reference weight < Lower weight limit × 10

0.1%

Lower weight limit

× 10 ≤ Reference weight < Lower weight limit × 100

0.01%

Lower weight limit

× 100 ≤ Reference weight

You cannot set a value below the lower weight limit as a reference weight.

* The lower weight limit depends on the model. For more information, refer to Section 23: Specifications

on page 83.