8 setting limit values, 9 sample input commands – Rice Lake TA Series, Rice Lake Weighing Systems User Manual

Page 71

Advertising
background image

RL-TA Series Operation Manual 67

17.4.8 Setting limit values

Command Main Body

Description

Value

Response

first

character

second

character

ASCII code

L

A

4CH

41H

First setting
point/lower limit

Limit

values

A00: Successful

completion

E01:Command error

E02:Set value error

L

B

4CH

42H

Second setting
point/upper limit

L

C

4CH

43H

Reference value

L

D

4CH

44H

Third setting point

L

E

4CH

45H

Forth setting point

1.

Insert a comma between command entries and limit values; e.g., LA, nnn.

2.

Enter numerical values only. You do not need to enter units.

3.

In weighing machine mode, the unit set in Unit A is applied to the entered value. For example, when

“LA, 20.00” is transmitted, it is recognized as 20.00 g if gram is selected in Unit A, and 20.00 ct if

carat is selected in Unit A.

4.

The units used for parts counting, percentage weighing, and unit converting are PCS, %, and #

respectively.

17.4.9 Sample input commands

T(SP)(CR)(LF)

Set tare range or adjust to zero-point.

O1(CR)(LF)

Set the balance to continuous output.

O8(CR)(LF)

Output data (once immediately).

IA,01,30,00(CR)(LF)

Set an interval of 01 hrs: 30 min: 00 sec.

OA(CR)(LF)

Start the interval function.

DD(CR)(LF)

Output the date.

DT(CR)(LF)

Output the time.

LA,80.5(CR)(LF)

Set the first setting point to 80.5. (The unit varies depending on the weighing

mode.)